GUM-compliant measurement uncertainty budget components and coverage factors for common engineering measurements. Covers Type A (statistical) and Type B (non-statistical) uncertainty sources with standard uncertainties, expanded uncertainties at k=2 (95%), and confidence levels for: length/dimension (caliper, micrometer, CMM, gauge block), mass (analytical balance, platform scale), temperature (Type K thermocouple, Pt100 RTD, IR pyrometer), pressure (Bourdon gauge, electronic transducer, liquid manometer), flow (orifice plate, turbine meter, Coriolis, ultrasonic transit-time), electrical (DMM, power analyzer), force (load cell), and torque (torque wrench). Also includes coverage factor reference rows (k=1/2/3, 68/95/99.7%) and distribution divisors (rectangular, triangular, normal, U-shaped). Per ISO/IEC Guide 98-3:2008 (GUM), NIST TN 1297, ISO/IEC 17025:2017.

General Referencemeasurement_type: pressure_bourdon_gauge3 rows
measurement typeuncertainty sourceconfidence level pct (%)coverage factor kdistributiondivisorexpanded uncertainty k2expanded uncertainty unitnotesstandard uncertainty unitstandard uncertainty valueuncertainty type
pressure_bourdon_gaugeaccuracy_class952rectangular1.7320.58pct_FSBourdon gauge ASME B40.100 Grade 2A: accuracy +/-0.5% FS; half-interval 0.25%; u = 0.25/sqrt(3) = 0.144% FS; Grade 1A: +/-1% FS; reported as combined 0.29% FSpct_FS0.29Type_B
pressure_bourdon_gaugecalibration952normal20.4pct_FSDeadweight tester calibration uncertainty; U95 = 0.05-0.1% FS for good deadweight tester; expanded to 0.20% FS for field gauge calibration labpct_FS0.2Type_B
pressure_bourdon_gaugecombined_standard_uncertainty952normal10.76pct_FSRSS: sqrt(0.29^2+0.14^2+0.20^2+0.10^2) = 0.38% FS; U95 = 0.76% FS; combined Bourdon gauge measurement uncertaintypct_FS0.38Combined