Measurement Uncertainty: flow_turbine_meter

GUM-compliant measurement uncertainty budget components and coverage factors for common engineering measurements. Covers Type A (statistical) and Type B (non-statistical) uncertainty sources with standard uncertainties, expanded uncertainties at k=2 (95%), and confidence levels for: length/dimension (caliper, micrometer, CMM, gauge block), mass (analytical balance, platform scale), temperature (Type K thermocouple, Pt100 RTD, IR pyrometer), pressure (Bourdon gauge, electronic transducer, liquid manometer), flow (orifice plate, turbine meter, Coriolis, ultrasonic transit-time), electrical (DMM, power analyzer), force (load cell), and torque (torque wrench). Also includes coverage factor reference rows (k=1/2/3, 68/95/99.7%) and distribution divisors (rectangular, triangular, normal, U-shaped). Per ISO/IEC Guide 98-3:2008 (GUM), NIST TN 1297, ISO/IEC 17025:2017.

General Referencemeasurement_type: flow_turbine_meter3 rows
measurement typeuncertainty sourceconfidence level pct (%)coverage factor kdistributiondivisorexpanded uncertainty k2expanded uncertainty unitnotesstandard uncertainty unitstandard uncertainty valueuncertainty type
flow_turbine_metercalibration_factor952normal20.5pct_flowTurbine meter K-factor calibration uncertainty; typical linearity +/-0.15% over 10:1 range; cal at flow lab U95 = 0.5%pct_flow0.25Type_B
flow_turbine_metercombined_standard_uncertainty952normal10.66pct_flowRSS: sqrt(0.25^2+0.20^2+0.05^2) = 0.323% flow; U95 = 0.65% flow; turbine meters among most accurate for clean low-viscosity fluidspct_flow0.33Combined
flow_turbine_meterrepeatability952normal10.1pct_flowTurbine meter repeatability typically 0.05-0.10% of reading; excellent short-term repeatabilitypct_flow0.05Type_A