Measurement Uncertainty: flow_coriolis_meter

GUM-compliant measurement uncertainty budget components and coverage factors for common engineering measurements. Covers Type A (statistical) and Type B (non-statistical) uncertainty sources with standard uncertainties, expanded uncertainties at k=2 (95%), and confidence levels for: length/dimension (caliper, micrometer, CMM, gauge block), mass (analytical balance, platform scale), temperature (Type K thermocouple, Pt100 RTD, IR pyrometer), pressure (Bourdon gauge, electronic transducer, liquid manometer), flow (orifice plate, turbine meter, Coriolis, ultrasonic transit-time), electrical (DMM, power analyzer), force (load cell), and torque (torque wrench). Also includes coverage factor reference rows (k=1/2/3, 68/95/99.7%) and distribution divisors (rectangular, triangular, normal, U-shaped). Per ISO/IEC Guide 98-3:2008 (GUM), NIST TN 1297, ISO/IEC 17025:2017.

General Referencemeasurement_type: flow_coriolis_meter3 rows
measurement typeuncertainty sourceconfidence level pct (%)coverage factor kdistributiondivisorexpanded uncertainty k2expanded uncertainty unitnotesstandard uncertainty unitstandard uncertainty valueuncertainty type
flow_coriolis_metercalibration952normal20.2pct_flowFactory calibration traceable to NIST; Coriolis accuracy class typically 0.1-0.2% of reading; expanded U95 = 0.20%pct_flow0.1Type_B
flow_coriolis_metercombined_standard_uncertainty952normal10.24pct_flowRSS: sqrt(0.05^2+0.10^2+0.03^2) = 0.116% flow; U95 = 0.23% flow; Coriolis is reference-class flow measurementpct_flow0.12Combined
flow_coriolis_meterrepeatability952normal10.06pct_flowExcellent repeatability: sd typically 0.02-0.05% for stable flow; one of best available flow technologiespct_flow0.03Type_A