Measurement Uncertainty: mass_platform_scale

GUM-compliant measurement uncertainty budget components and coverage factors for common engineering measurements. Covers Type A (statistical) and Type B (non-statistical) uncertainty sources with standard uncertainties, expanded uncertainties at k=2 (95%), and confidence levels for: length/dimension (caliper, micrometer, CMM, gauge block), mass (analytical balance, platform scale), temperature (Type K thermocouple, Pt100 RTD, IR pyrometer), pressure (Bourdon gauge, electronic transducer, liquid manometer), flow (orifice plate, turbine meter, Coriolis, ultrasonic transit-time), electrical (DMM, power analyzer), force (load cell), and torque (torque wrench). Also includes coverage factor reference rows (k=1/2/3, 68/95/99.7%) and distribution divisors (rectangular, triangular, normal, U-shaped). Per ISO/IEC Guide 98-3:2008 (GUM), NIST TN 1297, ISO/IEC 17025:2017.

General Referencemeasurement_type: mass_platform_scale3 rows
measurement typeuncertainty sourceconfidence level pct (%)coverage factor kdistributiondivisorexpanded uncertainty k2expanded uncertainty unitnotesstandard uncertainty unitstandard uncertainty valueuncertainty type
mass_platform_scalecalibration952normal20.2gCalibration uncertainty using Class F1 weights; typical expanded uncertainty U = 0.20 g at k=2 for 30 kg scale; includes eccentricity and spang0.1Type_B
mass_platform_scalecombined_standard_uncertainty952normal10.36gRSS: sqrt(0.15^2+0.10^2+0.05^2) = 0.185 g; U95 = 0.37 g for typical 30 kg platform scaleg0.18Combined
mass_platform_scalerepeatability952normal10.1gType A uncertainty from 10 repeated weighings; sd typically 0.05-0.10 g for 30 kg platform scaleg0.05Type_A

This is a sample. Search the full measurement uncertainty dataset and 300+ more by creating a free account and accessing the database.

Get started for FREE