Biocompatibility Ratings: surface_device#skin
ISO 10993-1:2018 biological evaluation requirements for medical devices. For each combination of device category (surface_device, external_communicating, implant), contact type (skin, mucosal_membrane, breached_surface, tissue_bone, blood_path_indirect, blood_contacting, tissue_bone_implant, blood_implant), and contact duration (limited_lt_24hr, prolonged_24hr_to_30d, permanent_gt_30d), specifies which biocompatibility tests are required per ISO 10993-1:2018 Table A.1 and FDA Guidance (September 2020). Test endpoints include cytotoxicity, sensitization, irritation, acute systemic toxicity, subacute/subchronic/chronic toxicity, genotoxicity, implantation, hemocompatibility, carcinogenicity, reproductive toxicity, and degradation. Supplementary note records document key footnotes and risk-based exceptions from ISO 10993-1:2018 Annex A.
| biocompat key | contact duration | acute systemic toxicity required | carcinogenicity required | chronic toxicity required | contact type | cytotoxicity required | degradation required | device category | genotoxicity required | hemocompatibility required | implantation required | irritation required | material examples | reproductive toxicity required | sensitization required | subacute toxicity required | subchronic toxicity required |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| surface_device#skin | limited_lt_24hr | no | no | no | skin | yes | no | surface_device | no | no | no | yes | surgical gloves, electrode pads, wound dressings (short-term) | no | yes | no | no |
| surface_device#skin | limited_lt_24hr_notes | no | no | no | skin | no | no | surface_device | no | no | no | no | Note: ISO 10993-1:2018 Table A.1 footnote — cytotoxicity and sensitization are the minimum for ALL surface devices; additional tests required based on risk assessment | no | no | no | no |
| surface_device#skin | permanent_gt_30d | yes | yes | yes | skin | yes | no | surface_device | yes | no | no | yes | ostomy adhesives, long-term transdermal patches, long-wear electrodes | no | yes | no | yes |