Biocompatibility Ratings: external_communicating#tissue_bone
ISO 10993-1:2018 biological evaluation requirements for medical devices. For each combination of device category (surface_device, external_communicating, implant), contact type (skin, mucosal_membrane, breached_surface, tissue_bone, blood_path_indirect, blood_contacting, tissue_bone_implant, blood_implant), and contact duration (limited_lt_24hr, prolonged_24hr_to_30d, permanent_gt_30d), specifies which biocompatibility tests are required per ISO 10993-1:2018 Table A.1 and FDA Guidance (September 2020). Test endpoints include cytotoxicity, sensitization, irritation, acute systemic toxicity, subacute/subchronic/chronic toxicity, genotoxicity, implantation, hemocompatibility, carcinogenicity, reproductive toxicity, and degradation. Supplementary note records document key footnotes and risk-based exceptions from ISO 10993-1:2018 Annex A.
| biocompat key | contact duration | acute systemic toxicity required | carcinogenicity required | chronic toxicity required | contact type | cytotoxicity required | degradation required | device category | genotoxicity required | hemocompatibility required | implantation required | irritation required | material examples | reproductive toxicity required | sensitization required | subacute toxicity required | subchronic toxicity required |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| external_communicating#tissue_bone | limited_lt_24hr | yes | no | no | tissue_bone | yes | no | external_communicating | no | no | no | yes | surgical probes, biopsy forceps, laparoscopic instruments (single use) | no | yes | no | no |
| external_communicating#tissue_bone | limited_lt_24hr_implantation_note | no | no | no | tissue_bone | no | no | external_communicating | no | no | no | no | Note: implantation testing (ISO 10993-6) is recommended when the device contacts viable tissue; brief contact may waive implantation per FDA risk-based justification | no | no | no | no |
| external_communicating#tissue_bone | permanent_gt_30d | yes | yes | yes | tissue_bone | yes | yes | external_communicating | yes | no | yes | yes | neurostimulator leads, cochlear implant arrays, chronic percutaneous transcutaneous connectors | yes | yes | yes | yes |