Biocompatibility Ratings: external_communicating#blood_contacting

ISO 10993-1:2018 biological evaluation requirements for medical devices. For each combination of device category (surface_device, external_communicating, implant), contact type (skin, mucosal_membrane, breached_surface, tissue_bone, blood_path_indirect, blood_contacting, tissue_bone_implant, blood_implant), and contact duration (limited_lt_24hr, prolonged_24hr_to_30d, permanent_gt_30d), specifies which biocompatibility tests are required per ISO 10993-1:2018 Table A.1 and FDA Guidance (September 2020). Test endpoints include cytotoxicity, sensitization, irritation, acute systemic toxicity, subacute/subchronic/chronic toxicity, genotoxicity, implantation, hemocompatibility, carcinogenicity, reproductive toxicity, and degradation. Supplementary note records document key footnotes and risk-based exceptions from ISO 10993-1:2018 Annex A.

Materials Sciencebiocompat_key: external_communicating#blood_contacting3 rows
biocompat keycontact durationacute systemic toxicity requiredcarcinogenicity requiredchronic toxicity requiredcontact typecytotoxicity requireddegradation requireddevice categorygenotoxicity requiredhemocompatibility requiredimplantation requiredirritation requiredmaterial examplesreproductive toxicity requiredsensitization requiredsubacute toxicity requiredsubchronic toxicity required
external_communicating#blood_contactinglimited_lt_24hryesnonoblood_contactingyesnoexternal_communicatingyesyesnoyesarterial catheters (single use), vascular access catheters, blood sample syringesnoyesnono
external_communicating#blood_contactinglimited_lt_24hr_carcinogenicity_notenononoblood_contactingnonoexternal_communicatingnonononoNote: carcinogenicity (ISO 10993-3) typically waived for limited-contact blood-path devices via genotoxicity negative result plus risk assessment; required if leachable profiles indicate carcinogenic potentialnononono
external_communicating#blood_contactingpermanent_gt_30dyesyesyesblood_contactingyesyesexternal_communicatingyesyesnoyesventricular assist device blood-contacting surfaces, chronic arteriovenous shuntsyesyesyesyes

This is a sample. Search the full biocompatibility ratings dataset and 300+ more by creating a free account and accessing the database.