Home/Reference Data/Biocompatibility Ratings/implant#tissue_bone_implant

Biocompatibility Ratings: implant#tissue_bone_implant

ISO 10993-1:2018 biological evaluation requirements for medical devices. For each combination of device category (surface_device, external_communicating, implant), contact type (skin, mucosal_membrane, breached_surface, tissue_bone, blood_path_indirect, blood_contacting, tissue_bone_implant, blood_implant), and contact duration (limited_lt_24hr, prolonged_24hr_to_30d, permanent_gt_30d), specifies which biocompatibility tests are required per ISO 10993-1:2018 Table A.1 and FDA Guidance (September 2020). Test endpoints include cytotoxicity, sensitization, irritation, acute systemic toxicity, subacute/subchronic/chronic toxicity, genotoxicity, implantation, hemocompatibility, carcinogenicity, reproductive toxicity, and degradation. Supplementary note records document key footnotes and risk-based exceptions from ISO 10993-1:2018 Annex A.

Materials Sciencebiocompat_key: implant#tissue_bone_implant3 rows
biocompat keycontact durationacute systemic toxicity requiredcarcinogenicity requiredchronic toxicity requiredcontact typecytotoxicity requireddegradation requireddevice categorygenotoxicity requiredhemocompatibility requiredimplantation requiredirritation requiredmaterial examplesreproductive toxicity requiredsensitization requiredsubacute toxicity requiredsubchronic toxicity required
implant#tissue_bone_implantlimited_lt_24hryesnonotissue_bone_implantyesnoimplantnonoyesyestemporary surgical sponges, short-term intraperitoneal lavage devicesnoyesnono
implant#tissue_bone_implantpermanent_gt_30dyesyesyestissue_bone_implantyesyesimplantyesnoyesyesorthopaedic bone plates, screws, hip stems, knee components, spinal cages, dental implantsyesyesyesyes
implant#tissue_bone_implantpermanent_gt_30d_degradation_notenononotissue_bone_implantnonoimplantnonononoNote: degradation evaluation (ISO 10993-9/13/14/15) required for resorbable polymers (PLA/PGA/PCL), bioresorbable ceramics (TCP/HA), and metals (Mg alloys) when leachables/degradation products are anticipatednononono