ISO 10993-1:2018 biological evaluation requirements for medical devices. For each combination of device category (surface_device, external_communicating, implant), contact type (skin, mucosal_membrane, breached_surface, tissue_bone, blood_path_indirect, blood_contacting, tissue_bone_implant, blood_implant), and contact duration (limited_lt_24hr, prolonged_24hr_to_30d, permanent_gt_30d), specifies which biocompatibility tests are required per ISO 10993-1:2018 Table A.1 and FDA Guidance (September 2020). Test endpoints include cytotoxicity, sensitization, irritation, acute systemic toxicity, subacute/subchronic/chronic toxicity, genotoxicity, implantation, hemocompatibility, carcinogenicity, reproductive toxicity, and degradation. Supplementary note records document key footnotes and risk-based exceptions from ISO 10993-1:2018 Annex A.

Materials Sciencebiocompat_key: external_communicating#blood_path_indirect3 rows
biocompat keycontact durationacute systemic toxicity requiredcarcinogenicity requiredchronic toxicity requiredcontact typecytotoxicity requireddegradation requireddevice categorygenotoxicity requiredhemocompatibility requiredimplantation requiredirritation requiredmaterial examplesreproductive toxicity requiredsensitization requiredsubacute toxicity requiredsubchronic toxicity required
external_communicating#blood_path_indirectlimited_lt_24hryesnonoblood_path_indirectyesnoexternal_communicatingnoyesnoyessolution administration sets, IV tubing, blood-collecting tubesnoyesnono
external_communicating#blood_path_indirectpermanent_gt_30dyesyesyesblood_path_indirectyesnoexternal_communicatingyesyesnoyespermanent central venous catheters, implantable ports, long-term haemodialysis componentsyesyesyesyes
external_communicating#blood_path_indirectprolonged_24hr_to_30dyesnonoblood_path_indirectyesnoexternal_communicatingyesyesnoyesIV infusion sets (multi-day), enteral feeding tubes, short-term haemodialysis linesnoyesyesno