Sterilization Parameters: ultraviolet

Sterilization process parameters for medical devices, pharmaceutical equipment, and healthcare facilities. Covers ethylene oxide (EtO standard/low-temp/hospital cartridge), gamma irradiation (VDmax15, VDmax25, substantiation 25 kGy, high dose 35 kGy), electron beam, X-ray, steam autoclave (gravity 121°C, prevacuum 134°C, liquid cycle, terminal, flash, prion), hydrogen peroxide plasma/vapor (VPHP, VHP, accelerated), dry heat (150°C/170°C/HVHA), low-temperature steam formaldehyde (LTSF), ozone, UV-C, and cold plasma. Parameters include temperature, exposure time, radiation dose, gas concentration, relative humidity, sterility assurance level (SAL), material compatibility, and validation standard references.

Biomedical Engineeringsterilization_method: ultraviolet1 row
sterilization methodprocess variantcycle pressure conditionexposure time max min (min)exposure time min min (min)exposure time typical min (min)material compatibility notesmethod descriptionnotestemperature max C (°C)temperature min C (°C)temperature typical C (°C)validation standard
ultravioletuvc_surface3-4_log_reduction_surface_only601030atmospheric_ambientUV-C germicidal irradiation (254 nm) — surface decontamination only; NOT a sterilization modalityCDC and EPA guidance for surface disinfection; ASHRAE 185.1 (HVAC UV)402025Surface/air disinfection only; no penetration through materials; shadows prevent uniform treatment; used for cleanroom surfaces, water disinfection, UV-C cabinets for small items; does NOT achieve sterility for medical devices