Sterilization Parameters

Sterilization process parameters for medical devices, pharmaceutical equipment, and healthcare facilities. Covers ethylene oxide (EtO standard/low-temp/hospital cartridge), gamma irradiation (VDmax15, VDmax25, substantiation 25 kGy, high dose 35 kGy), electron beam, X-ray, steam autoclave (gravity 121°C, prevacuum 134°C, liquid cycle, terminal, flash, prion), hydrogen peroxide plasma/vapor (VPHP, VHP, accelerated), dry heat (150°C/170°C/HVHA), low-temperature steam formaldehyde (LTSF), ozone, UV-C, and cold plasma. Parameters include temperature, exposure time, radiation dose, gas concentration, relative humidity, sterility assurance level (SAL), material compatibility, and validation standard references.

Biomedical Engineering10 sections21 sample rowsKey: sterilization_method

Attributes

method_description
temperature_min_C°C
temperature_typical_C°C
temperature_max_C°C
exposure_time_min_minmin
exposure_time_typical_minmin
exposure_time_max_minmin
dose_kGy_minkGy
dose_kGy_typicalkGy
dose_kGy_maxkGy
gas_concentration_mg_per_L_minmg/L
gas_concentration_mg_per_L_typicalmg/L
gas_concentration_mg_per_L_maxmg/L
relative_humidity_pct_min%
relative_humidity_pct_typical%
relative_humidity_pct_max%
SAL
cycle_pressure_condition
material_compatibility_notes
validation_standard
notes