Sterilization Parameters: ozone

Sterilization process parameters for medical devices, pharmaceutical equipment, and healthcare facilities. Covers ethylene oxide (EtO standard/low-temp/hospital cartridge), gamma irradiation (VDmax15, VDmax25, substantiation 25 kGy, high dose 35 kGy), electron beam, X-ray, steam autoclave (gravity 121°C, prevacuum 134°C, liquid cycle, terminal, flash, prion), hydrogen peroxide plasma/vapor (VPHP, VHP, accelerated), dry heat (150°C/170°C/HVHA), low-temperature steam formaldehyde (LTSF), ozone, UV-C, and cold plasma. Parameters include temperature, exposure time, radiation dose, gas concentration, relative humidity, sterility assurance level (SAL), material compatibility, and validation standard references.

Biomedical Engineeringsterilization_method: ozone1 row
sterilization methodprocess variantcycle pressure conditionexposure time max min (min)exposure time min min (min)exposure time typical min (min)material compatibility notesmethod descriptionnotestemperature max C (°C)temperature min C (°C)temperature typical C (°C)validation standard
ozoneozone_sterilization6-log18060120sub-atmosphericOzone gas sterilization (aqueous-phase and gaseous) for heat-sensitive devicesFDA 510(k) K040312 cleared (TSO3 SciCan Trios)30420Compatible with stainless steel, titanium, most plastics; incompatible with silicone rubber, natural rubber, some adhesives; no toxic residuals — breaks down to O2; limited penetration into lumens