Sterilization Parameters: hydrogen_peroxide_plasma

Sterilization process parameters for medical devices, pharmaceutical equipment, and healthcare facilities. Covers ethylene oxide (EtO standard/low-temp/hospital cartridge), gamma irradiation (VDmax15, VDmax25, substantiation 25 kGy, high dose 35 kGy), electron beam, X-ray, steam autoclave (gravity 121°C, prevacuum 134°C, liquid cycle, terminal, flash, prion), hydrogen peroxide plasma/vapor (VPHP, VHP, accelerated), dry heat (150°C/170°C/HVHA), low-temperature steam formaldehyde (LTSF), ozone, UV-C, and cold plasma. Parameters include temperature, exposure time, radiation dose, gas concentration, relative humidity, sterility assurance level (SAL), material compatibility, and validation standard references.

Biomedical Engineeringsterilization_method: hydrogen_peroxide_plasma3 rows
sterilization methodprocess variantcycle pressure conditionexposure time max min (min)exposure time min min (min)exposure time typical min (min)material compatibility notesmethod descriptionnotestemperature max C (°C)temperature min C (°C)temperature typical C (°C)validation standard
hydrogen_peroxide_plasmaaccelerated_h2o21e-6552840sub-atmospheric_1-10_TorrAccelerated H2O2 sterilization — Sterizone/TSO3 advanced H2O2 system with multiple injectionFDA 510(k); CE mark503545Improved lumen access vs standard VPHP; multi-pulse injection; biological indicator G. stearothermophilus; validated for longer/narrower lumens than Sterrad
hydrogen_peroxide_plasmavhp_vaporized6-log_bio_decontamination480120240sub-atmospheric_or_near-atmVaporized hydrogen peroxide without plasma activation — room decontamination and larger chamberEPA/CDC for room decontamination; Bioquell/STERIS VHP402030Larger chambers possible without plasma; used for room decontamination, isolators, cleanroom decon; not classified as sterilizer by FDA — decontamination standard
hydrogen_peroxide_plasmavphp_low_temperature1e-6752855sub-atmospheric_1-7_TorrVaporized/Plasma Hydrogen Peroxide (VH2O2/VPHP) sterilization at low temperatureFDA 510(k) cleared; EU MDR 2017/745553745Excellent for heat/moisture sensitive devices: endoscopes, electronics, batteries, metals, most plastics; not suitable for cellulosic materials (absorbs H2O2), long narrow lumens, liquids

This is a sample. Search the full sterilization parameters dataset and 300+ more by creating a free account and accessing the database.