Prosthetics Orthotics Design: upper_limb_prosthesis

Structural design loads, socket interface pressures, alignment tolerances, and component specifications for prosthetic limbs and orthotic devices. Covers transtibial (below-knee) and transfemoral (above-knee) prostheses, KAFO/AFO orthotics, and upper-limb prostheses. ISO 10328 cyclic load levels P3/P4/P5/P6, K-level activity classifications K0-K4, peak ground reaction forces, socket interface pressures by tissue type, and alignment tolerances. Per ISO 10328:2016, ISO 22523:2006, Frossard mechanical loading studies, CMS K-level definitions.

Biomedical Engineeringdevice_type: upper_limb_prosthesis2 rows
device typedesign parameterconditiondescriptionnotesunitvalue max
upper_limb_prosthesisstructural_load_transhumeralstatic_functional_testISO 22523 structural load for transhumeral (above-elbow) — P4750 N at wrist level for AE prosthesis. Above-elbow prosthesis: elbow joint rated for suspension loads.N750
upper_limb_prosthesisstructural_load_transradialstatic_functional_testISO 10328-equivalent load for transradial (below-elbow) — P4ISO 22523 load per limb category. 500 N applies to TD-elbow junction for transradial prosthesis P4 class.N500