Structural design loads, socket interface pressures, alignment tolerances, and component specifications for prosthetic limbs and orthotic devices. Covers transtibial (below-knee) and transfemoral (above-knee) prostheses, KAFO/AFO orthotics, and upper-limb prostheses. ISO 10328 cyclic load levels P3/P4/P5/P6, K-level activity classifications K0-K4, peak ground reaction forces, socket interface pressures by tissue type, and alignment tolerances. Per ISO 10328:2016, ISO 22523:2006, Frossard mechanical loading studies, CMS K-level definitions.
| device type | design parameter | condition | description | notes | unit | value max | value min | value typical |
|---|---|---|---|---|---|---|---|---|
| transfemoral_prosthesis | alignment_socket_flexion | design_spec | AK socket flexion — initial socket flexion angle | AK socket set 0-5 deg flexion (minimal) to provide stability in stance. High-activity users may use more. | degrees | 15 | 0 | 5 |
| transfemoral_prosthesis | knee_joint_load_peak | walking_midstance | Peak load through prosthetic knee joint — level walking | Knee joint axial load. Varies with activity, cadence, body mass. 80-kg user produces ~1.0-1.4 kN through stance. | N | 1,600 | 1,000 | 1,200 |