Explore the Engineering Database more efficiently by signing up for free. Sign up free
Resolution-Noise — Pressure Transducer
- Measurement Type
- Pressure Transducer
- Uncertainty Source
- Resolution Noise
- Confidence Level Pct (%)
- 95
- Coverage Factor K
- 2
- Distribution
- Rectangular
- Divisor
- 1.732
- Expanded Uncertainty K2
- 0.06
- Expanded Uncertainty Unit
- Pct Fs
- Standard
- ASME B40.100; Instrument Manufacturer Spec
- Standard Uncertainty Unit
- Pct Fs
- Standard Uncertainty Value
- 0.03
- Uncertainty Type
- Type B
Source: ASME B40.100; instrument manufacturer spec