Explore the Engineering Database more efficiently by signing up for free. Sign up free

Calibration — Pressure Bourdon Gauge

Measurement Type
Pressure Bourdon Gauge
Uncertainty Source
Calibration
Confidence Level Pct (%)
95
Coverage Factor K
2
Distribution
Normal
Divisor
2
Expanded Uncertainty K2
0.4
Expanded Uncertainty Unit
Pct Fs
Standard
ASME B40.100; Nist Deadweight Tester Calibration
Standard Uncertainty Unit
Pct Fs
Standard Uncertainty Value
0.2
Uncertainty Type
Type B

Source: ASME B40.100; NIST deadweight tester calibration