Buoyancy-Correction — Mass Analytical Balance
- Measurement Type
- Mass Analytical Balance
- Uncertainty Source
- Buoyancy Correction
- Confidence Level Pct (%)
- 95
- Coverage Factor K
- 2
- Distribution
- Normal
- Divisor
- 2
- Expanded Uncertainty K2
- 0.002
- Expanded Uncertainty Unit
- G
- Standard
- Oiml R 111-1; Euramet Cg-18 V4 Section 5
- Standard Uncertainty Unit
- G
- Standard Uncertainty Value
- 0.001
- Uncertainty Type
- Type B
Source: OIML R 111-1; EURAMET cg-18 v4 Section 5