Explore the Engineering Database more efficiently by signing up for free. Sign up free

Buoyancy-Correction — Mass Analytical Balance

Measurement Type
Mass Analytical Balance
Uncertainty Source
Buoyancy Correction
Confidence Level Pct (%)
95
Coverage Factor K
2
Distribution
Normal
Divisor
2
Expanded Uncertainty K2
0.002
Expanded Uncertainty Unit
G
Standard
Oiml R 111-1; Euramet Cg-18 V4 Section 5
Standard Uncertainty Unit
G
Standard Uncertainty Value
0.001
Uncertainty Type
Type B

Source: OIML R 111-1; EURAMET cg-18 v4 Section 5