Temperature-Stability — Length Cmm

Measurement Type
Length Cmm
Uncertainty Source
Temperature Stability
Confidence Level Pct (%)
95
Coverage Factor K
2
Distribution
Rectangular
Divisor
1.732
Expanded Uncertainty K2
0.004
Expanded Uncertainty Unit
MM
Standard
ISO 10360-2:2009 Annex A; Manufacturer Environmental Specs
Standard Uncertainty Unit
MM
Standard Uncertainty Value
0.002
Uncertainty Type
Type B

Source: ISO 10360-2:2009 Annex A; manufacturer environmental specs