Temperature-Stability — Length Cmm
- Measurement Type
- Length Cmm
- Uncertainty Source
- Temperature Stability
- Confidence Level Pct (%)
- 95
- Coverage Factor K
- 2
- Distribution
- Rectangular
- Divisor
- 1.732
- Expanded Uncertainty K2
- 0.004
- Expanded Uncertainty Unit
- MM
- Standard
- ISO 10360-2:2009 Annex A; Manufacturer Environmental Specs
- Standard Uncertainty Unit
- MM
- Standard Uncertainty Value
- 0.002
- Uncertainty Type
- Type B
Source: ISO 10360-2:2009 Annex A; manufacturer environmental specs