Explore the Engineering Database more efficiently by signing up for free. Sign up free
Repeatability — Length Cmm
- Measurement Type
- Length Cmm
- Uncertainty Source
- Repeatability
- Confidence Level Pct (%)
- 95
- Coverage Factor K
- 2
- Distribution
- Normal
- Divisor
- 1
- Expanded Uncertainty K2
- 0.002
- Expanded Uncertainty Unit
- MM
- Standard
- ISO 10360-2:2009 Section 5; ASME B89.4.10360
- Standard Uncertainty Unit
- MM
- Standard Uncertainty Value
- 0.001
- Uncertainty Type
- Type a
Source: ISO 10360-2:2009 Section 5; ASME B89.4.10360