Temperature-Coefficient — Length Caliper
- Measurement Type
- Length Caliper
- Uncertainty Source
- Temperature Coefficient
- Confidence Level Pct (%)
- 95
- Coverage Factor K
- 2
- Distribution
- Rectangular
- Divisor
- 1.732
- Expanded Uncertainty K2
- 0.008
- Expanded Uncertainty Unit
- MM
- Standard
- ISO 1:2022 Standard Reference Temperature 20C; Nist Tn 1297 Section 8
- Standard Uncertainty Unit
- MM
- Standard Uncertainty Value
- 0.004
- Uncertainty Type
- Type B
Source: ISO 1:2022 standard reference temperature 20C; NIST TN 1297 Section 8