Explore the Engineering Database more efficiently by signing up for free. Sign up free

Thermal-Emf — Electrical Multimeter Dmm

Measurement Type
Electrical Multimeter Dmm
Uncertainty Source
Thermal Emf
Confidence Level Pct (%)
95
Coverage Factor K
2
Distribution
Normal
Divisor
2
Expanded Uncertainty K2
0.02
Expanded Uncertainty Unit
MV
Standard
IEEE 1376-1996; Fluke 87V Specification
Standard Uncertainty Unit
MV
Standard Uncertainty Value
0.01
Uncertainty Type
Type B

Source: IEEE 1376-1996; Fluke 87V specification

electrical_multimeter_dmm — uncertainty source thermal_emf · standard IEEE 1376-1996; Fluke 87V specification · standard uncertainty value 0.01 · expanded uncertainty k2 0.02 · expanded uncertainty unit mV | Measurement Uncertainty | Engineering Database