Explore the Engineering Database more efficiently by signing up for free. Sign up free
Calibration — Electrical Multimeter Dmm
- Measurement Type
- Electrical Multimeter Dmm
- Uncertainty Source
- Calibration
- Confidence Level Pct (%)
- 95
- Coverage Factor K
- 2
- Distribution
- Normal
- Divisor
- 2
- Expanded Uncertainty K2
- 0.1
- Expanded Uncertainty Unit
- Pct Rdg
- Standard
- Fluke 87V specification; ANSI/ISA-51.1-1979
- Standard Uncertainty Unit
- Pct Rdg
- Standard Uncertainty Value
- 0.05
- Uncertainty Type
- Type B
Source: Fluke 87V specification; ANSI/ISA-51.1-1979