Explore the Engineering Database more efficiently by signing up for free. Sign up free
Self-Heating — Temperature RTD PT100
- Measurement Type
- Temperature RTD PT100
- Uncertainty Source
- Self Heating
- Confidence Level Pct (%)
- 95
- Coverage Factor K
- 2
- Distribution
- Normal
- Divisor
- 2
- Expanded Uncertainty K2
- 0.06
- Expanded Uncertainty Unit
- degC
- Standard
- IEC 60751:2008 Section 4.3; Instrument Manufacturer Spec
- Standard Uncertainty Unit
- degC
- Standard Uncertainty Value
- 0.03
- Uncertainty Type
- Type B
Source: IEC 60751:2008 Section 4.3; instrument manufacturer spec